>> On a tour of the range Micron ion exchanger M2
The MICRON column uses a nuclear quality mixed bed, ionically balanced, macroporous.It is placed between the ion exchangers and the terminal filter and gives the water a final polishing. The great porosity of its resin allows the 0.1 micron particles, including the colloidal silica, to be removed. By retention it removes the slightly ionised organic compounds. Conductivity with 1 R2 upstream : 0,3 to 2 µS/cm. Flow rate : 20 liters/hour Pressure : 2,5 bar (at ambient temperature) pH (at 20°C) : 5,5 à 7 Maximum use temperature : 60°C Saturation : saturation of the MICRON column is not visible by discolouration. They should be replaced as soon as colloidal particles reappear - this can only be revealed by analisys. However, for averagely polluted water, this replacement will be done after the replacement of six R2 columns. |